Alarms
156
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
CPP
Increment:
1 mmHg
or 0.1 kPa
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
100 mmHg
(13.3 kPa)
70 mmHg
(9.3 kPa)
Off
Tblood
Increment of
0.1
°C
or 0.1 °F
Upper: 25.1 to 43.0 °C
77.1 to 109.4 °F
Lower: 25.0 to 42.9 °C
77.0 to 109.2 °F
39.0 °C
(102.2 °F)
34.0 °C
(93.2 °F)
Off
FiO2
Increment of 1%
On
Upper: 19 to 100%
Lower: 18 to 99%
100%
20%
Store
etO
2
Increment of 1%
Upper: 11 to 100%
Lower: 10 to 99%
100%
10%
Off
inN
2
O
On
(fixed)
Fixed at 82%
82%
Not applicable
Store
(fixed)
RRc
Increment of
1
/min
(Adult/
pediatric)
On
(neonate)
Upper: 6 to 150 /min
(when no CO
2
device is
connected)
6 to 100 /min (when Scio
is connected)
Lower: 5 to 149 /min
(when no CO
2
device is
connected)
5 to 99 /min (when Scio is
connected)
NOTE
Scio is not available in
neonate mode.
30 /min (adult)
60 /min
(pediatric/
neonate)
5 /min (adult)
20 /min
(pediatric/
neonate)
Off
Parameter Alarm
default
status
Alarm limit range
Upper limit
defaults
Lower limit
defaults
Archive
default
setting