Impedance respiratory rate (RRi)
248
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Accessing the respiration settings
Select the ventilation parameter field to select
the
Resp.
page directly.
or
1
Select
Sensor parameters...
from the main
menu bar.
2
Select the
Resp.
tab to access the
Resp.
page.
If you cannot see the tab, select the following
two symbols located in the upper right corner of
the dialog:
symbol and the display filter
button.
3
Select the
Settings 1
and
Settings 2
tabs.
The top portion of the
Settings 1
page contains
the
Auto set
and
Alarm
buttons for configuring
the alarm functions. For detailed alarm setup
information, see "Configuring the alarm settings
for a patient" on page 125.
Ventilation parameter setup functions
All respiration setup functions take place in the
Resp.
page.
Selection
Available settings
Description
Settings 1
page
Resp. lead
I, II (default)
Selects the lead for respiration
monitoring.
Relearn
None
Initiates a relearning of the
respiration signal.
Mode
Auto
(default),
Manual
(see page 247 for
more details).
Determines the processing
mode for the breath-related
impedance change.
Size [%]
10% to 100% (in 10% increments) –
default:
50%
Adjusts the waveform size
and/or breath detection
threshold, according to the
selected respiration setting.
–
Auto
mode – Waveform
size only, without affecting
the breath-detection
threshold.
–
Manual
mode – Waveform
size
and
breath-detection
threshold.