Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
147
Alarms
External device disconnection alarm
If the external device alarm feature is activated at
the Cockpit (see page 461) and an external device
is disconnected from the Cockpit, the following
happens at the ICS and at the Cockpit:
– An alarm tone of low priority sounds.
– The message
External device disconnected
appears.
The
Code
function
You can configure a set of monitoring functions that
can be activated during emergency care by
selecting the
Code
button on the main menu bar.
Depending on which of these settings are activated
(see page 464), any of the following happens when
you select the
Code
button:
– A continuous recording starts
– NIBP measurements start in continuous mode
– The alarm volume of the alarm condition with
the highest priority is automatically reduced to
the minimum setting.
– The pre-configured
All alarms off
setting
determines if the alarm annunciation is
deactivated and the
All alarms off
message
appears in the header bar when you press the
Code
button (see page 465).
In addition to activating the pre-configured features,
a timer with a red background appears in the
header bar with the following two buttons:
–
Stop
for stopping the timer. The label of the
button changes to
Start
.
–
Reset
button for resetting the timer to zero.
The
Code
button does not function unless the
M540 is docked.
To activate the Code function
Press the
Code
button on the main menu bar.
To deactivate the Code function
Press the
Code
button on the main menu bar a
second time. All functions are deactivated.
NOTE
When the
Audio off
setting is set to
Off
(see
page 461), the message
Audio off
and the
symbol
appear in the Cockpit header bar
when you invoke the Code function.
NOTE
The
All alarms off
button works independently of
the pre-configured
Code
setting. You can
therefore activate or deactivate the
All alarms off
button even after pressing the
Code
key.
NOTE
When the
All alarms off
setting is set to
Off
(see
page 461), the message
Audio alarms off
and
the
symbol appear in the Cockpit header bar
when you invoke the Code function.