![Dräger Infinity Acute Care System M540 Instructions For Use Manual Download Page 172](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773172.webp)
Trends/data dialogs
172
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Selecting an interval
The interval that applies to the three graphical
trends can be set.
To set the interval
1
Select the
Duration
button (F).
2
Choose one of the following durations:
Off
,
1 min
,
5 min
,
10 min
(default),
15 min
,
20 min
,
30 min
,
45 min
,
1 h
,
90
min
,
2 h
,
4
h.
3
Select the
OK
button (C) to confirm the
selection. Select the
Cancel
button (D) to exit
the screen without accepting the changes.
Printing an Analysis tool graphical trend
report
A graphical trend report for recruitment contains the
initial and end cursor values and the
Δ
value for
each parameter. graphical trend report for
recruitment can be printed only after you set the
cursors. Otherwise the
button remains grayed
out and cannot be selected.
To print a recruitment graphical trend report
1
Select the
Procedures...
button from the main
menu bar.
2
Select the
Analysis tool
tab.
3
Set both cursor buttons (see page 171).
4
Select the
button.
Tabular trend
The tabular trend displays trend data in data
columns. Trend data are updated according to the
selected time scale. For example, if the current time
scale is 15 minutes, the trend display is updated
every 15 minutes. A time stamp above each
column marks the interval during which the data in
that column was collected. The displayed value is
the last acquired value during that interval. The
column on the right side is reserved for the most
recent data. Certain parameters and special
conditions, such as artifact, are represented in
unique ways (see page 162).
To access the tabular trend
1
Select the
Trends/ Data...
button on the main
menu bar.
2
Select the
Trends
tab (if not already selected).
3
Select the
Table
tab (if not already selected).