Technical data
592
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Infinity R50N
Physical attributes
Size (H x W x D)
180 x 120 x 222 mm
(7.1 x 4.7 x 8.7 in)
Weight
1.6 kg (3.6 lbs)
Connections
AC power
X14 Infinity network connector
Potential equalization connector
Cooling
Convection
Power requirements
Input voltage
100 to 240 V AC (50/60 Hz, 1 A)
Risk management
Protection class
Class 1
Protection against liquid ingress
IPX0 according to IEC 60529.
Mode of operation
Continuous
Environmental requirements
Operation
Temperature
15 to 40 °C (59 to 104 °F)
Relative humidity
10 to 95%, non-condensing
Ambient pressure
550 to 775 mmHg (73 to 103 kPa)
Storage
Temperature range
–20 to +40 °C (–4 to +104 °F)
Relative humidity
10 to 95%, non-condensing without packaging
Ambient pressure
375 to 795 mmHg (50 to 106 kPa)