Sensor and Cable Compliance
WARNING:
The use of accessories, sensors, and cables other than those specified may
result in inaccurate readings of the monitoring system and increased emission
of the monitoring system.
Table 2-9.
Sensor and Cable Length
Item
SKU
Maximum Length
Sensors
Nellcor™ Adult SpO
2
Sensor, Reusable (Nonsterile)
DS100A
3.0 ft. (0.9 m)
Nellcor™ Adult XL SpO
2
Sensor (Sterile, single-use only)
MAX-AL
3.0 ft. (0.9 m)
Nellcor™ Forehead SpO
2
Sensor (Sterile, single-use only)
MAX-FAST
2.5 ft (0.75 m)
Nellcor™ Neonatal-Adult SpO
2
Sensor
(Sterile, single-use only)
MAX-N
1.5 ft. (0.5 m)
Nellcor™ Infant SpO
2
Sensor (Sterile, single-use only)
MAX-I
Nellcor™ Pediatric SpO
2
Sensor (Sterile, single-use only)
MAX-P
Nellcor™ Adult SpO
2
Sensor (Sterile, single-use only)
MAX-A
Nellcor™ Adult SpO
2
Nasal Sensor (Sterile, single-use only)
MAX-R
Nellcor™ Adult-Neonatal SpO
2
Sensor with Wraps
(Reusable with adhesive)
OXI-A/N
3.0 ft. (0.9 m)
Nellcor™ Pediatric-Infant SpO
2
Sensor with Wraps
(Reusable with adhesive)
OXI-P/I
Nellcor™ Pediatric SpO
2
Sensor, Two Piece
(Sterile, single-use only)
P
OC-3 cable,
3.0 ft. (0.9 m)
Nellcor™ Neonatal-Adult SpO
2
Sensor, Two Piece
(Sterile, single-use only)
N
Nellcor™ Adult SpO
2
Sensor, Two Piece
(Sterile, single-use only)
A
Nellcor™ SpO
2
Sensor, Multisite Reusable (Nonsterile)
D-YS
4.0 ft. (1.2 m)
•
Nellcor™ SpO
2
Ear Clip, Reusable (Nonsterile)
D-YSE
•
Nellcor™ Pediatric SpO
2
Clip, Reusable (Nonsterile)
D-YSPD
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......