Modification and Testing
10-44
Service Manual
10.8
Monitoring Screen Calibration
Only calibrate the monitoring screen if it stops responding appropriately to human
touch. If caregivers report problems accessing menu items and the screen is
unlocked, consider calibration. Calibration requires a stylus or very fine, rounded
tip of a non-marking utensil. Do not use the pointed tip of a pencil or pen.
To calibrate the monitoring screen
1.
Access the service function menu. Reference
To access service function menus
2.
Select the SERVICE menu.
3.
Select the SERVICE option.
4.
Select TOUCH SCREEN CALIBRATION.
5.
Press YES.
6.
Press NEXT.
7.
Press and hold stylus in exact center of crosshatch in center of monitoring screen.
Figure 10-19.
Initial Calibration Screen
8.
Press and hold stylus in exact center of crosshatch as it moves to each corner of
the monitoring screen.
9.
Insert a USB keyboard in the USB port.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......