![background image](http://html1.mh-extra.com/html/covidien/nellcor-gr101704/nellcor-gr101704_service-manual_2673898082.webp)
Operation
6-20
Service Manual
Alarm System
, p. 10-6, for modification of alarm
system defaults.
Alarm Limits
Use this to set institutional alarm limit thresholds.
•
Set permissions for parameter alarms —
Use to set SatSeconds, Pulse Rate
Delay, and SPD parameter permissions.
•
Set permissions for alarm limit and audio alarm defaults —
Use to set per-
missions for alarm limit adjustments, alarm audio OFF, setting alarm silence dura-
tion, and setting the alarm disabled reminder.
•
Set desired alarm silence duration default —
Use to select the desired alarm
silence duration default.
•
Set desired alarm limit thresholds for adult and neonate mode —
Use to
set SpO
2
and pulse rate alarm limit thresholds, as well as SatSeconds, Pulse Rate
Delay, and SPD parameter default settings.
To set SatSeconds™, Pulse Rate Delay, and SPD™ Alert parameter defaults
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select ALARM SYSTEM.
4.
Select the appropriate parameter.
a.
Select ALLOW SATSECONDS to set the SatSeconds permissions default.
b.
Select ALLOW PULSE RATE DELAY to set the Pulse Rate Delay permissions
default.
c.
Select ALLOW SPD to set the SPD permissions default.
d.
Select SPD AUDIO ALERT to set the audio alert default for the SPD parameter.
5.
Select the desired default.
a.
Select NO if the institution does not permit caregivers using the parameter.
b.
Leave it YES if the institution permits caregivers using the parameter.
6.
Select the desired option.
a.
Press SAVE CHANGES to retain the change.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......