Modification and Testing
10-38
Service Manual
Figure 10-17.
LIGHT Test: BPM 60, SpO
2
90, MOD low, Light High
a.
No active visual and audio alarms.
b.
Saturation indication between 88 and 92 inclusive.
c.
BPM indication between 57 and 63.
d.
Pulse waveform of approximately half-inch P-T-P amplitude. Actual amplitude
may vary but references low pulse amplitude/high light indications.
15.
Press %SpO
2
.
16.
Verify the 75% LED lights. The monitoring system registers SpO
2
decreases and
stabilizes to a value between 73 and 77 inclusive.
a.
Active visual and audio alarms.
b.
Saturation indication between 73 and 77 inclusive, creating an SpO
2
LOW
alarm condition.
c.
BPM indication between 57 and 63 inclusive.
d.
Pulse waveform of approximately half-inch peak to peak (P-T-P) amplitude.
Actual amplitude may vary but will be a reference for low pulse amplitude/
high light indications.
17.
Press MOD.
18.
Verify the MOD increase LED lights. The pulse amplitude waveform initially increases in
amplitude and then stabilize at P-T-P amplitude of approximately one-inch.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......