Modification and Testing
10-20
Service Manual
Note:
If the alarm duration set to OFF, the display immediately shows AUDIO OFF.
5.
Adjust the alarm volume setting to an appropriate audio level for the environment
in which it is to operate.
6.
Press SAVE CHANGES.
To configure pulse beep volume
Adjust the pulse tone volume after adjusting the alarm volume. Reference
1.
Turn on the monitoring system by pressing the POWER ON key.
2.
Press MENU.
3.
Press SOUND SETTINGS.
4.
Select PULSE BEEP VOLUME.The monitoring system displays an volume icon with
increment and decrement buttons.
5.
Press decrement (-) until the pulse beep volume setting displays 25%. Verify the
volume of the alarm decreases.
6.
Press increment (+) to increase the pulse beep volume setting to a maximum value
of 100%. Verify the volume increases.
7.
Adjust the pulse beep volume setting to an appropriate audio level for the envi-
ronment in which it is to operate.
8.
Press SAVE CHANGES.
To configure button click volume
Adjust the pulse tone volume after adjusting the alarm volume. Reference
1.
Turn on the monitoring system by pressing the POWER ON key.
2.
Press MENU.
3.
Press SOUND SETTINGS.
4.
Select BUTTON CLICK VOLUME.The monitoring system displays an volume icon
with increment and decrement buttons.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......