Service Mode
Service Manual
6-25
To set the default main trend scale or monitoring history trend scale
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select DISPLAY SETTINGS.
4.
Select the desired option.
a.
MAIN SCREEN TREND SCALE
b.
MONITORING HISTORY TREND SCALE
5.
Select the desired default.
a.
Minutes: 15 or 30
b.
Hours: 1, 2, 4, 8, 12, 24, or 48
6.
Select the desired option.
a.
Press SAVE CHANGES to retain the change.
b.
Press CANCEL to leave the default as it was.
To adjust default backlight brightness
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select DISPLAY SETTINGS.
4.
Select BACKLIGHT BRIGHTNESS.
5.
Adjust to the desired default brightness.
a.
Press decrement (-) to decrease the default brightness.
b.
Press increment (+) to increase the default brightness.
6.
Select the desired option.
a.
Press SAVE CHANGES to retain the change.
b.
Press CANCEL to leave the default as it was.
To set backlight OFF permission
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......