Performance Tests
Service Manual
10-17
8.
Ensure the FAST icon appears on the monitoring screen next to %SpO
2
.
To test alarm silence duration
1.
Turn on the monitoring system by pressing the POWER ON key.
2.
Press MENU.
3.
Press MONITORING SETTINGS.
4.
Select ALARM SILENCE DURATION.
5.
Ensure the mode is set to 120 seconds.
6.
Select 30 seconds.
7.
Press SAVE CHANGES.
8.
Press SILENCE ALARM.
9.
Ensure the silenced alarm icon appears above both saturation and pulse rate alarm
threshold settings.
10.
Ensure the silence timer counts down from 0:30 to 0:00.
10.6.2
Operational Setup
Operational Setup procedures confirm settings for and allow configuration of
the listed parameters.
•
To configure alarms and alarm silence,
•
•
To configure pulse beep volume,
•
To configure button click volume,
•
To calibrate analog voltage output,
To configure alarms and alarm silence
1.
Connect the DOC-10 extension cable to the sensor port.
2.
Connect the Nellcor™ Adult SpO
2
Sensor (DS-100A) to the interface cable, then
to a finger.
3.
Turn on the monitoring system by pressing the POWER ON key.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......