Service Mode
Service Manual
6-29
To set default Nurse Call polarity
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select CONNECTIVITY SETTINGS.
4.
Select NURSE CALL POLARITY.
5.
Select the desired default: Normally high (+) or normally low (-).
6.
Select the desired option.
a.
Press SAVE CHANGES to retain the change.
b.
Press CANCEL to leave the default as it was.
Monitoring Settings
•
Alarm Mode —
Use to set the default alarm mode to adult settings or neonate
settings.
•
Date and Time —
Use to set the default display for date and time or after a soft-
ware update.
•
Response Mode —
Use to set the default response mode.
To set alarm mode default
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select MONITORING SETTINGS.
4.
Select ALARM MODE.
5.
Select the desired default.
a.
Adult alarm settings
b.
Neonate alarm settings
6.
Select the desired option.
a.
Press SAVE CHANGES to retain the change.
b.
Press CANCEL to leave the default as it was.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......