Product Specifications
2-12
Service Manual
2.8.2
Ground Integrity
100 milliohms or less
2.8.3
Safety Tests
The following tables describe the maximum earth and enclosure leakage
current allowed, as well as patient leakage.
Cables
Power cord
----
9.84 ft. (3 m)
DOC-10 interface cable
10.0 ft. (3 m)
Firmware download cable, RS-232 serial, 15 to 9 pin “D”
10.0 ft. (3 m)
Non-terminated cable, RS-232 analog, 15 pin “D”
3.3 ft. (1 m)
Printer cable, RS-232, 15 to 9 pin “D”
10.0 ft. (3 m)
Philips interface cable
M1943 NL
3.3 ft. (1 m)
Oxinet™ III hardwire cable
----
10.0 ft. (3 m)
Oxinet™ III data cable
Table 2-9.
Sensor and Cable Length (Continued)
Item
SKU
Maximum Length
Table 2-10.
Earth and Enclosure Leakage Current Specifications
Earth Leakage Current
Condition
AC Line Polarity
Line Cord
Neutral
Line Cord
IEC 60601-1
ANSI/AAMI
60601-1
Normal
Normal
Closed
Closed
500 μA
300 μA
Single Fault
Open
Closed
1000 μA
Closed
Open
Normal
Reversed
Closed
Closed
500 μA
300 μA
Single Fault
Open
Closed
1000 μA
Closed
Open
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......