Trend Data Access
7-14
Service Manual
6.
Disable the Dynamic Host Configuration Protocol by pressing the button next to
“Obtain an IP address automatically.”
7.
Attach a USB keyboard.
8.
Input the IP address and subnet mask.
9.
Scroll to the bottom of the screen and press OK.
10.
Close the window by pressing the “X” in the top, right corner.
11.
Detach the USB keyboard.
12.
Press FINISH.
7.3.4
Data Port Communications
To setup data port communication
1.
While in normal monitoring mode, press MENU.
2.
Press CONNECTIVITY SETTINGS.
3.
Select the desired protocol for data exchange.
4.
Press SAVE CHANGES to save the selected setting.
7.4
Using the Nurse Call Interface
7.4.1
Nurse Call Feature
WARNING:
Do not use the nurse call feature as the primary source of alarm notification.
The audible and visual alarms of the monitoring system, used in conjunction
with clinical signs and symptoms, are the primary sources for notifying
medical personnel that an alarm condition exists.
WARNING:
The nurse call feature of the monitoring system is operational when the
monitoring system is powered by AC power or battery power. However, the
nurse call feature does not function when monitoring system alarms are
silenced.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......