Troubleshooting
11-8
Service Manual
Sample alarm condition
The monitoring system may detect a failure of the primary speaker and sound a
high-pitched piezo tone. A primary speaker failure alarm appears.
To access an alarm message
1.
Press MORE INFO or VIEW ALL to continue, depending on which is available. A
description of the error and any recommended action appears. This particular
alarm cannot be cleared.
Figure 11-4.
Sample Speaker Failure Message
2.
Press ALARM SILENCE to silence the high-pitched piezo tone. This provides the
caregiver two (2) minutes to resolve the issue.
3.
Take the recommended action to resolve the issue.