Repair
12-36
Service Manual
Figure 12-20.
Antennae PCB and UFL Connectors Removal
To replace the Antennae PCBs
1.
Gently thread wires through from the front to the rear of the Main PCB, with the
ferrite bead remaining visible and each antenna PCB slipping through the left or
right opening on the Main PCB.
2.
Gently connect both golden UFL antennae connectors onto the SBC PCB.
3.
Wire tie ferrite bead to the Main PCB, capturing right antenna PCB wire under-
neath the wire tie, while ensuring the wire remains free under the Main PCB.
4.
Slide the left-hand antenna PCB under the supporting clip and onto its standouts.
5.
Slip right-hand antenna PCB back in under the plastic LCD clamp plate.
6.
Replace the Main PCB. Reference
7.
Reassemble the monitoring system chassis. Reference
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......