Repair
12-8
Service Manual
2.
Remove and set aside the battery cover.
3.
Slide out the battery while pulling on the battery tab.
Figure 12-4.
Battery Removal
To replace the battery access door and battery
1.
After removing the old battery, gently slide the new battery into the battery slot,
contact side down and facing the slot.
2.
Replace the battery cover, hooking the tab into the back chassis.
3.
Place and tighten the T-15 torx battery cover screw to 10 inch-pounds.
4.
Connect to AC power.
5.
Press the ON/STANDBY button to power on the monitoring system.
6.
Verify battery charge and AC power indicators illuminate.
7.
Allow the battery to fully charge prior to any clinical use.
8.
Verify unit functionality. Reference
12.4.3
Rubber Feet Replacement
To remove rubber feet
1.
Gently rotate the monitoring system, placing so the bottom of the chassis is up.
2.
Slide the tip of the flat-blade screwdriver under each rubber foot and pry up to remove.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......