Modification and Testing
10-16
Service Manual
e.
Select PULSE RATE DELAY.
f.
Ensure the delay is set to OFF.
8.
Turn off the monitoring system by pressing the POWER ON key.
Testing Monitoring Settings
Service technicians may choose to test all monitoring settings in sequence. If
so, reserve turning off the monitoring system until the end of the sequence. If
not, turn it off after completing the individual test to clear all changes.
To test alarm mode
1.
Turn on the monitoring system by pressing the POWER ON key.
2.
Press MENU.
3.
Press MONITORING SETTINGS.
4.
Select ALARM MODE.
5.
Ensure the alarm mode is ADULT ALARM SETTINGS.
6.
Select NEONATE ALARM SETTINGS.
7.
Press SAVE CHANGES.
8.
Ensure the Baby icon appears on the main monitoring screen between both satu-
ration and pulse rate alarm threshold settings.
9.
Press YES.
To test response mode
1.
Turn on the monitoring system by pressing the POWER ON key.
2.
Press MENU.
3.
Press MONITORING SETTINGS.
4.
Select RESPONSE MODE.
5.
Ensure the response mode is NORMAL.
6.
Select FAST.
7.
Press SAVE CHANGES.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......