Performance Tests
Service Manual
10-23
9.
Repeat Step 8.
10.
Verify the intensity continues to decrease. This variation is an indication the micro-
processor is in proper control of LED intensity.
11.
Turn off the monitoring system by pressing the POWER ON key.
Operation with a Live Subject
Patient monitoring involves connecting the DS-100A sensor to a live subject for
a qualitative test.
To test using a live subject:
1.
Ensure the monitoring system is connected to an AC power source.
2.
Connect a DOC-10 interface cable to the sensor port.
3.
Connect the DS-100A sensor to the interface cable.
4.
Attach the sensor to a live subject as recommended in the sensor's
Instructions For
Use
.
5.
Turn on the monitoring system by pressing the POWER ON key.
6.
Verify the monitoring system properly identifies the attached sensor. The monitor-
ing system should stabilize on the subject's physiological signal in approximately
15 to 30 seconds.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......