Alarm Issues
Service Manual
11-15
The Pulse Search indicator illu-
minates after successful mea-
surements occur.
•
Check the status of the patient.
•
Perfusion may be too low for the monitoring system to track the pulse.
Test the monitoring system on another patient. Change the sensor site
and/or try another type of Nellcor™ sensor.
•
Interference may be preventing the monitoring system from tracking
the pulse. Verify the sensor is securely applied and replace it if neces-
sary. Change the sensor site. Use a type of sensor that tolerates more
patient movement; for example, an Nellcor™ adhesive sensor. Electro-
magnetic interference may be preventing the monitoring system from
tracking the pulse. Remove the source of interference and/or try to
stabilize the environment.
•
The sensor may be too tight, there may be excessive ambient light, or
the sensor may be on an extremity with a blood pressure cuff, arterial
catheter, or intravascular line. Reposition the sensor as necessary.
Alarms only briefly actuate.
•
Check patient trend data.
•
Examine the patient.
•
Determine if Alarm Audio Off is enabled. If so, an alarm will sound
every three (3) minutes.
•
Determine if the SPD alarm is active and alarming.
•
Check the speaker. Reference
Table 11-6.
Alarm Issues (Continued)
Problem
Resolution
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......