Power
Service Manual
6-3
Battery Fuel Gauge
Caution:
Should a low battery alarm sound, connect the monitoring system to an AC
power source and then silence the alarm by pressing ALARM SILENCE. If the
monitoring system is operated on an AC power source with a depleted
battery and AC power is subsequently lost, the monitoring system will shut
down immediately.
The monitoring system runs on an internal battery when not connected to an
AC power source. A battery fuel gauge displays the remaining battery power.
When connected to AC power, the battery fuel gauge displays a lightning bolt
while charging and at full charge.
Connection to an AC Power Source
Note:
The battery is recyclable. Do not dispose of the battery by placing it in the regular
trash. Dispose of the battery in accordance with local guidelines and regulations or
contact Covidien to arrange for disposal.
Note:
As the battery is used and recharged over time, the amount of time between the onset
of low battery alarms and the monitoring system shut-off may become shorter.
Table 6-1.
Battery Power Status
1
1. The levels listed are based on a new battery. Continued battery charge and discharge eventually reduces capacity. For
example, a battery two years old may provide only 75% of the capacity of a new battery.
Message
Color
Power Charge Status
None
Green
Normal Status —
Indicates 15-100% (approximately 15
minutes to 6 hours) battery capacity remains.
BATTERY LOW
Yellow
Low Status —
Indicates 5-14% (approximately 15 minutes)
battery capacity remains.
BATTERY CRITICALLY
LOW
Red
Critical Status —
Indicates 1-4% (approximately 5 minutes)
battery capacity remains.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......