Front Panel Components Replacement
Service Manual
12-37
Single Board Computer (SBC) PCB Replacement
To remove the Single Board Computer (SBC) PCB
1.
Disassemble the monitoring system chassis. Reference
2.
Move the rear enclosure off to the side, retaining the front panel on a static-free
working surface.
3.
Remove and set aside the four T-6 torx screws.
4.
Gently disconnect both golden UFL antennae connectors for the antennae from
the board.
5.
Gently remove the SBC board, exerting even pressure on both sides.
Figure 12-21.
Single Board Computer (SBC) Removal
To replace the Single Board Computer (SBC) PCB
1.
Align the Single Board Computer (SBC) PCB connectors to the staggered connec-
tors on the Main PCB.
2.
Gently press down to connect the SBC PCB to the Main PCB, ensuring the board
is completely seated.
3.
Reconnect both golden UFL antennae connectors.
4.
Tighten the four T-6 torx screws to 3 inch-pounds.
5.
Realign the rear enclosure to the front panel on a static-free working surface.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......