Product Specifications
2-2
Service Manual
2.3.2
Battery
Note:
The battery provides approximately seven hours of battery life when new and fully-charged
with no alarms, no serial data, no analog output, no nurse call output, with backlight on
while using a pulse simulator set for 200 bpm, high light and low modulation.
2.3.3
Rating of Nurse Call Relay
Type
Lithium Ion
Voltage
7.2 Volts DC, 11.6 Ah, 83 Wh
Recharge
8 hours with monitoring system turned off
12 hours with monitoring system turned on
Shelf Life
Four months, if monitoring system runs on new, fully-charged battery
After four months storage, units run 33% of stated battery life
Compliance
IEC 62133
Maximum Input Voltage
30 VAC or VDC (polarity is not important)
Load Current
120 mA continuous (peak 300 mA @ 100 ms)
Minimum Resistance
26.5 ohms to 50.5 ohms (40.5 ohms typical) during alarms
Ground Reference
Isolated Ground
Electrical Isolation
1500 Volts
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......