Repair
12-18
Service Manual
To disassemble the monitoring system chassis
1.
Turn off by pressing the ON/STANDBY button.
2.
Disconnect from the AC power source.
3.
Remove the parameter module. Reference
To remove the parameter module
11.
4.
Set it face down on a static-free work surface.
Caution:
Ensure the work surface is clean and free of debris.
5.
To remove the battery access door and battery
6.
Remove and set aside all four T-15 torx screws, one from each corner of the rear
chassis.
Figure 12-11.
Corner Chassis Screws Removal
7.
Carefully separate front and back chassis cases without stressing the connection
harnesses between them.
8.
Gently set the back chassis on its feet.
Caution:
To avoid stress or damage to the wiring harness, ensure the front chassis
is lying flat on the work surface.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......