Troubleshooting
11-12
Service Manual
Unlit LEDs on the power switch
membrane.
•
Check for power connectivity.
•
Check for cable damage.
•
Replace the front chassis.
Powers down or resets without
apparent cause
•
Remove from power and allow to cool down. After it is cool, recon-
nect to power. If it continues to reset, check the fan. If the fan is func-
tional, remove to a controlled environment and retry. If it still resets or
overheats, replace the SBC PCB.
•
After confirming it is not a hardware issue, contact Technical Services.
BATTERY MISSING message
occurs, even though a battery is
present.
•
The battery charger system is not functioning properly. Replace the
battery with a known good battery. If this resolves the problem,
replace the defective battery.
•
If the problem persists with a known good battery, the battery charger
system is not functioning properly. Replace the Main PCB.
•
If this does not resolve the issue, return both the battery and the mon-
itoring system to Covidien Technical Services.
Table 11-4.
Power Failure Issues (Continued)
Problem
Resolution
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......