Modification and Testing
10-28
Service Manual
Figure 10-7.
BPM Test: BPM 200 and SpO
2
75
a.
Active visual and audio alarms.
b.
Saturation indication between 73 and 77 inclusive, creating an SpO
2
LOW
alarm condition.
c.
BPM indication between 194 and 206, creating a PULSE RATE HIGH alarm
condition.
d.
Pulse waveform of approximately 1/2-inch P-T-P amplitude. Actual amplitude
may vary but references low pulse amplitude/low light patients.
9.
Press PULSE RATE.
10.
Verify the 60 BPM LED lights. The monitoring system registers pulse rate decreases
and stabilizes to a value between 57 to 63 BPM inclusive.
a.
Active visual and audio alarms.
b.
Saturation indication between 73 and 77 inclusive, creating an SpO
2
LOW
alarm condition.
c.
BPM indication between 57 and 63 inclusive.
d.
Pulse waveform of approximately 1/2-inch peak to peak (P-T-P) amplitude.
Actual amplitude may vary but will be a reference for low pulse amplitude/low
light patients.
11.
Turn off the monitoring system by pressing the POWER ON key.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......