Repair
12-24
Service Manual
4.
Repeat this process for the remaining ground wire cable ring terminal.
5.
Firmly connect the green (GND) cable to the single prong to the top of the power
entry module.
6.
Firmly connect the blue (N) cable to the left-hand, neutral (N) blade of the power
entry module and the brown (L) cable to the right-hand, live (L) blade of the power
entry module.
7.
Firmly connect the green (GND) spade connector to the J21 mating blade on the
Main PCB.
8.
Firmly connect the green (GND) spade connector to the E1 mating blade on the
power supply PCB.
9.
Firmly connect two-pin blue (N) and brown (L) (AC) cable and connector to the
2-pin J1 receptacle on the power supply PCB.
10.
Reassemble the monitoring system chassis. Reference
12.6.3
Battery Components Replacement
The battery cradle assembly consists of several subassemblies and PCBs. For suc-
cessful battery cradle replacement, perform the following instructions in order.
To remove and then replace battery components
1.
Remove the battery. Reference
To remove the battery access door and battery
2.
Disassemble the chassis. Reference
Monitoring System Chassis Disassembly
3.
Remove the battery cradle. Reference
Removing the Battery Cradle and Compo-
4.
Remove and replace the Power Supply PCB. Reference
5.
Remove and replace the Battery Interconnect PCB. Reference
6.
Remove and replace the cooling fan. Reference
7.
Replace the battery cradle. Reference
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......