Power Components Replacement
Service Manual
12-25
8.
Reverse the process for replacement.
Battery Cradle Replacement
Removing the Battery Cradle and Components
1.
Disassemble the monitoring system chassis. Reference
2.
Move the front chassis off to the side.
3.
Remove all four T-10 torx screws, two on top, and two deep underneath, using a
long-reach driver.
Figure 12-14.
Battery Cradle Removal
4.
Lift out the battery cradle, supporting both the battery cradle and still connected wires.
Note:
Cooling fan is still attached by wires, but is free-floating.
5.
Set it on a static-free work surface.
Caution:
Ensure the work surface is clean and free of debris.
6.
Remove the Power Supply PCB. Reference
7.
Remove the Battery Interconnect PCB. Reference