Chassis Disassembly and Reassembly
Service Manual
12-17
Caution:
Ensure the work surface is clean and free of debris.
3.
Orient the oximetry module PCB so the 8-pin J5 connector pins mate to the
parameter board J9 receptacle and the 10-pin J4 connector pins mate to the
parameter board J7 receptacle.
4.
Press downward on the oximetry module PCB until fully seated on the parameter
(interface) board PCB.
5.
Replace and tighten the two T-10 torx screws to a maximum of 6 inch-pounds.
6.
Rotate the parameter board assembly so the oximetry module PCB is facing down-
ward and both connector sockets are facing the front of the bottom enclosure.
7.
Place on the standouts to the left of the bottom enclosure.
8.
Replace and tighten the four T-10 torx screws to a maximum of 10 inch-pounds.
9.
Replace the top cover, ensuring the tabs slip down into the mating slots of the
bottom enclosure.
10.
Replace the faceplate, aligning to the mating tabs of the bottom enclosure and
ensuring the lugs of the top cover slip under the faceplate lip.
11.
Replace the parameter module. Reference
To replace the parameter module
12.
Connect to AC power.
13.
Verify unit functionality. Reference
12.5.2
Monitoring System Chassis Disassembly
Caution:
Observe ESD (electrostatic discharge) precautions when disassembling and
reassembling the monitoring system and when handling any internal
components.
Note:
The battery charge procedure should be performed before monitoring system repairs
when possible.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......