Front Panel Components Replacement
Service Manual
12-39
12.7.2
LCD Assembly with Overlay Replacement
WARNING:
The LCD panel contains toxic chemicals. Do not touch broken LCD panels.
Physical contact with a broken LCD panel can result in transmission or
ingestion of toxic substances.
To remove the LCD Assembly
1.
Disassemble the monitoring system chassis. Reference
2.
Move the rear enclosure off to the side, retaining the front panel on a static-free
working surface.
3.
Follow the steps for disconnecting the speaker. Reference
4.
Follow the steps for disconnecting both antennae PCBs. Reference
5.
Follow the steps for removing the Main PCB, setting it aside on a static-free
working surface. Reference
6.
Remove the three T-15 torx screws from the plastic LCD clamp plate.
7.
Gently lift and set aside the plastic LCD clamp plate, ensuring all flex cables remain free.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......