Repair
12-28
Service Manual
7.
Place the fan back down in its enclosure.
8.
Replace the battery cradle in the back chassis. Reference
9.
Reassemble the monitoring system chassis. Reference
Battery Interconnect PCB Replacement
To remove the Battery Interconnect PCB
1.
Follow all previous steps for removing the battery cradle. Reference
, p. 12-25. It is not necessary to remove the power supply PCB.
2.
Disconnect the five-pin J2 connector from the recess between the power supply
PCB and the wire-tied cable assembly with the ferrite bead.
3.
Rotate the battery cradle to ensure battery interconnect PCB accessibility.
4.
Remove both T-10 torx screws.
5.
Lift from the battery cradle.