Chassis Disassembly and Reassembly
Service Manual
12-15
5.
Replace the top cover, ensuring the tabs slip down into the mating slots of the
bottom enclosure.
6.
Replace the faceplate, aligning to the mating tabs of the bottom enclosure and
ensuring the lugs of the top cover slip under the faceplate lip.
7.
Orient and slide the new parameter module into the rear chassis until it snaps into place.
8.
Replace and tighten the T-15 torx screw to a maximum 10 inch-pounds.
9.
Connect to AC power.
10.
Verify unit functionality. Reference
Parameter Board PCBA Replacements
The parameter board assembly consists of two PCBs: the parameter (interface)
board PCB and the oximetry module PCB.
To remove the parameter board assembly
1.
Turn off the monitoring system by pressing the ON/STANDBY button.
2.
Disconnect from AC power.
3.
Remove the parameter module. Reference
To remove the parameter module
4.
Set it down on a static-free work surface.
Caution:
Ensure the work surface is clean and free of debris.
5.
Rotate the module so the slot in the side panel is visible.
6.
Using a flat-blade screwdriver, press inward on the visible faceplate tab by press-
ing inward, listening for a click to indicate tab release.
7.
Rotate the module so the slot in the opposite side panel is visible and repeat the
process.
8.
Remove and set aside the faceplate.
9.
Gently lift outward and upward on the top cover, releasing it from the mating
slots in the bottom enclosure.
10.
Remove and set aside the top cover.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......