Modification and Testing
10-18
Service Manual
4.
Press MENU.
5.
Select ALARM LIMITS.
6.
Select SpO
2
.
7.
Slide the SpO
2
Lower Alarm Limit up until it indicates 99. This automatically also
sets the SpO
2
Upper Alarm Limit to 100.
8.
Press SAVE CHANGES.
9.
Press MENU.
10.
Select ALARM LIMITS.
11.
Select PULSE RATE.
12.
Slide the Pulse Rate Lower Alarm Limit up until it indicates 160.
13.
Confirm the following results:
a.
The waveform tracks the pulse rate.
b.
The Pulse Tone is audible.
c.
The SpO
2
and pulse alarms are flashing yellow.
d.
Both audible alarms sound and visual alarms appear, indicating both parame-
ters have violated alarm limits.
Figure 10-2.
Sample: Configuring Alarms