Power Components Replacement
Service Manual
12-29
Figure 12-16.
Battery Interconnect PCB Removal
To replace the Battery Interconnect PCB
1.
Place the new battery interconnect PCB on the battery cradle.
2.
Replace and tighten both T-10 torx screws to 10 inch-pounds.
3.
Rotate the battery cradle.
4.
Reconnect the five-pin J2 connector.
5.
Replace the cooling fan. Reference
6.
Replace the battery cradle in the back chassis. Reference
Cooling Fan Replacement
To remove the cooling fan
1.
Follow all previous steps for removing the battery cradle. It is not necessary to
remove the power supply PCB or battery interconnect PCB.