Troubleshooting
11-18
Service Manual
Monitor was reset unexpectedly.
•
If problem persists, remove from service.
•
Check cabling and connectivity. Replace if needed.
•
Check fan. The monitoring system may be overheating.
Replace if needed.
•
Check the Main PCB. Replace if needed.
•
Check the SBC PCB. Replace if needed.
•
Check the power supply. Replace if needed.
SpO
2
or pulse rate data has not been
updated for over 30 seconds.
•
Check cabling and connectivity.
•
Nurse call output system error.
Serial Communications system failure.
Analog output system failure.
Cooling fan failure.
•
Check fan. Replace if needed.
•
Check the Main PCB. Replace if needed.
Thermal control system is above thermal
limit.
•
Power down and allow to cool.
•
Remove any potential sources of heat.
•
Check heat and humidity conditions. Reference
•
Move to a controlled environment.
•
Check fan. Replace if needed.
•
Check the Main PCB. Replace if needed.
Invalid date and time value detected during
start up.
•
Set the date and time.
•
Replace the battery, if stored for too long.
•
Check the Main PCB. Replace if needed.
BATTERY MISSING message occurs, even
though a battery is present.
Table 11-9.
Common Prompts and Error Messages (Continued)
Problem
Resolution
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......