Operation
6-16
Service Manual
6.4.6
HELP Option
To access on-screen help topics
1.
Press HELP. The appropriate help dialog window appears.
2.
Review the help dialog box for guidance.
3.
Press EXIT to return to normal monitoring.
6.5
Service Mode
Table 6-2.
Possible User Interface Settings
Option
Possible Setting
Factory Default
Display Settings
Monitor layout view
Pleth, Trend, Blip, Pleth and trend
Pleth
Main screen trend parameters
SpO
2
, Pulse, SpO
2
and pulse
SpO
2
and pulse
Main screen trend scale
15 or 30 minutes, 1, 2, 4, 8, 12, 24, or 48 hours
1 hour
Monitoring history trend scale
15 or 30 minutes, 1, 2, 4, 8, 12, 24, or 48 hours
1 hour
Screen Brightness
0, 25, 50, 75, 100%
75%
Allow backlight OFF
Yes, No
No
Wake display on alarm
Yes, No
Yes
Language
Croatian, Czech, Danish, Dutch, English, Estonian,
Finnish, French, German, Greek, Hungarian, Italian,
Japanese, Lithuanian, Norwegian, Polish, Portu-
guese, Romanian, Russian, Serbian, Simplified Chi-
nese, Slovak, Slovenian, Spanish, Swedish, Turkish
English
Sound Settings
Alarm Volume
0, 25, 50, 75, 100%
(min. 45dB, max > 85dB)
75%
Button Click Volume
50%
Pulse Beep Volume
50%
Monitoring Settings
Date and time
Time: 24 hour format, Date: and date format
24 hour, MM/DD/YY
Response mode
Normal, Fast
Normal
Alarm mode
Adult/Neonate
Adult
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......