Manufacturer’s Declaration and Guidance
Service Manual
2-7
Electromagnetic Emissions
Table 2-5.
Electromagnetic Emissions Guidelines and Compliance
Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
(IEC/EN 60601-1-2:2007, Table 1)
The monitoring system is intended for use in the electromagnetic environment specified below. The
customer or the user of the monitoring system should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment Guidance
RF emission
CISPR 11
EN 55011
Group 1,
Class A
Not intended for use in a residential environment. If
used in a domestic environment, may not offer ade-
quate protection to radio-frequency communication
services. The user may be required to take mitigation
measures, such as relocating or re-orienting the equip-
ment.
Harmonic emissions
IEC/EN 61000-3-2
Class A
N/A
Voltage fluctuation/
flicker emissions
IEC/EN 61000-3-3
Complies
N/A
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......