Modification and Testing
10-32
Service Manual
Figure 10-11.
MOD Test: BPM 60, SpO
2
75, and MOD High
a.
Active visual and audio alarms.
b.
Saturation indication between 73 and 77 inclusive.
c.
BPM indication between 57 and 63 inclusive.
d.
Pulse waveform of approximately one-inch P-T-P amplitude. Actual amplitude
may vary but references
high
pulse amplitude/low light patients.
9.
Press PULSE RATE.
10.
Verify the 200 BPM LED lights. The monitoring system registers pulse rate increas-
es and stabilizes to a value between 194 to 206 BPM inclusive.