Repair
12-16
Service Manual
11.
Remove and set aside the four T-10 torx parameter (interface) board PCB screws.
12.
Gently rotate the parameter board assembly, laying the assembly flat on a static-
free work surface with the oximetry module PCB facing up.
13.
Remove and set aside the two oximetry module PCB T-10 torx screws.
Figure 12-10.
Parameter Board PCB and Oximetry Module Removal
14.
Gently lift upward, removing the oximetry module PCB off the parameter board
assembly standoffs, ensuring all pins fully disconnect from both connectors.
15.
Remove and set aside the failing PCB in an alternate location.
To replace the parameter board assembly
1.
Obtain the replacement parameter board assembly component: the oximetry
module PCB or the parameter board PCB.
2.
Set with the known good component on a static-free work surface.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......