Performance Tests
Service Manual
10-11
6.
Disconnect the monitoring system from AC power.
7.
Verify the AC Power indicator is not lit.
To charge the battery
1.
Connect the monitoring system to an AC power source.
2.
Verify the AC Power indicator is lit.
3.
If the Low Battery indicator is also lit, leave the monitoring system connected to
an AC power source to charge the battery.
4.
Check the battery charge.
10.6
Performance Tests
The monitoring system automatically sets the Alarm Mode for adult alarm set-
tings at power on. Do not change this selection to neonate alarm settings for
any of these tests.
Note:
This section uses factory defaults. If the institution or caregiver using the monitoring
system customized the defaults, the monitoring system displays those customized
values. Reference
10.6.1
Power-On Defaults and Alarm Ranges
Power-up performance tests verify both power-on self-test (POST) and power-
on defaults and alarm range limits. Reference
Note:
Power-on defaults are factory settings or defaults set by the institution.
Note:
The descriptions below are based on the default pleth view with SatSeconds and the
Saturation Pattern Detection (SPD) feature enabled. The steps for changing an alarm
limit are the same for all views.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......