Modification and Testing
10-36
Service Manual
d.
Pulse waveform peak to peak (P-T-P) amplitude of approximately half-inch.
Actual amplitude may vary but will be a reference for low pulse amplitude/low
light patients.
7.
Press LIGHT LEVEL.
8.
Verify the increase LED lights. The pulse amplitude waveform initially flatlines and
then stabilizes at the same P-T-P amplitude of approximately half-inch.
Note:
Flat-lining is the only indication of a light change at the measurement site. If the
monitoring system recovers and displays normally, this is an indication of proper
operation with light changes.
Figure 10-15.
LIGHT Test: BPM 60, SpO
2
75, MOD low, Light High
a.
Active visual and audio alarms.
b.
Saturation indication between 73 and 77 inclusive.
c.
BPM indication between 57 and 63 inclusive.
d.
Pulse waveform of approximately half-inch P-T-P amplitude. Actual amplitude
may vary but references low pulse amplitude/
high
light indications.
9.
Press PULSE RATE.
10.
Verify the 200 BPM LED lights. The monitoring system registers pulse rate increas-
es and stabilizes to a value between 194 to 206 BPM inclusive.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......