Trend Data Access
7-16
Service Manual
create an alarm condition (for example, SENSOR DISCONNECT) and verify acti-
vation of the institution's nurse call system.
7.4.2
Setting Nurse Call RS-232 Polarity
The nurse call polarity can be set to a positive signal or a negative signal during
an alarm condition.
To set nurse call polarity
1.
While in normal monitoring mode, press MENU.
2.
Press CONNECTIVITY SETTINGS.
3.
Press NURSE CALL SETTINGS.
Figure 7-7.
Nurse Call Polarity Screen
4.
Press NORM + for a normally high OR press NORM - for a normally low setting.
5.
Press SAVE CHANGES to save the selected setting.