Modification and Testing
10-10
Service Manual
10.5
Battery Check
Replace batteries every 24 months or when battery charging rapidly depletes.
10.5.1
Battery Power
To check battery power
1.
Disconnect the monitoring system from AC power.
2.
Verify the AC Power indicator is not lit.
3.
Connect the monitoring system to an AC power source.
4.
Verify the AC Power indicator is lit.
5.
If the Low Battery indicator is also lit, leave the monitoring system connected to
an AC power source to charge the battery.
10.5.2
Battery Charge
Ensure the battery is fully charged by inspecting the Battery Fuel Gauge Indica-
tor, which should be completely full, and check the Low Battery Indicator, which
should be off. If the fuel gauge is not completely full, or if the Low Battery Indi-
cator is lit, charge the battery by connecting it to AC power. Reference
Note:
The battery charge procedure should be performed before monitoring system repairs
when possible.
To check battery charge
1.
Connect the monitoring system to an AC power source.
2.
Verify the AC Power indicator is lit.
3.
Allow a complete charge of the monitoring system.
4.
Turn on the monitoring system by pressing the POWER ON key.
5.
After POST completes, examine the battery fuel gauge for the percentage of
charge.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......