Alarms and Error Conditions
Service Manual
11-5
Alarm Prioritization
Alarms have an assigned priority. Stacked alarms display in order of criticality
and priority. Reference
, p. 6-11, for details on high, medium,
and low priority alarms. Occasionally, the monitoring screen field contains one
or more prompts or error messages. Stacked alarms display in order of critical-
ity and priority. High priority messages appear above low priority messages.
Alarms of the same priority appear in order of occurrence. If the caregiver does
not resolve the issue and clear the condition, the monitoring system may esca-
late the alarm by increasing the frequency of the alarm.
Table 11-2.
Initial Alarm Priority for Errors
Priority
Message
Condition
Resolution
Dismiss
Med.
Pulse Rate
Low
1.
If Pulse Rate Delay is enabled and not set
to OFF, patient pulse rate violates the
lower pulse rate limit threshold and
exceeds the Pulse Rate Delay limit
2.
If Pulse Rate Delay is disabled or set to
OFF, patient pulse rate violates the lower
pulse rate limit threshold
Pulse rate is below the alarm
limit. Check patient immedi-
ately.
No
Med.
Pulse Rate
High
1.
If Pulse Rate Delay is enabled and not set
to OFF, patient pulse rate violates the
upper pulse rate limit threshold and
exceeds the Pulse Rate Delay limit
2.
If Pulse Rate Delay is disabled or set to
OFF, patient pulse rate violates the upper
pulse rate limit threshold
Pulse rate is above the alarm
limit. Check patient immedi-
ately.
No
Med.
SpO
2
Low
1.
If the SatSeconds alarm is enabled and
not set to OFF, patient SpO
2
violates the
lower SpO
2
limit threshold AND exceeds
the SatSeconds limit
2.
If the SatSeconds alarm is disabled or set
to OFF, patient SpO
2
value violates the
lower SpO
2
limit threshold
SpO
2
is below the alarm
limit. Check patient immedi-
ately.
No
Med.
SpO
2
High
1.
If the SatSeconds alarm is enabled and
not set to OFF, patient SpO
2
violates the
upper SpO
2
limit threshold AND exceeds
the SatSeconds limit
2.
If the SatSeconds alarm is disabled or set
to OFF, patient SpO
2
value violates the
upper SpO
2
limit threshold
SpO
2
is above the alarm
limit. Check patient immedi-
ately.
No
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......