Performance Tests
Service Manual
10-19
14.
Press ALARM SILENCE.
15.
With the alarm silenced, verify the following:
a.
Alarm remains silenced for two (2) minutes.
b.
Alarm Silence indicator lights and time remaining counts down
c.
SpO
2
and BPM alarms continue to flash yellow
d.
SpO
2
and BPM values are surrounded by yellow
e.
Pulse tone is audible
f.
Audible alarm returns in approximately two (2) minutes
Figure 10-3.
Sample: Configuring Alarms with Silenced Alarm
To configure alarm volume
Configure the desired volume level of the alarm, from a silent 0% to a loud
100%. The factory default volume level is set to 75%.
1.
Turn on the monitoring system by pressing the POWER ON key.
2.
Press MENU.
3.
Select SOUND SETTINGS.
4.
Select ALARM VOLUME.The monitoring system displays an volume icon with
increment and decrement buttons.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......