Operation
6-18
Service Manual
All service function menus are accessible when the DOC-10 interface cable is
disconnected from the monitoring system. Disconnect the DOC-10 interface
cable from the monitoring system or disconnect the sensor from the DOC-10
interface cable.
To access service function menus
1.
Connect the monitoring system to an AC power source.
2.
Disconnect any Nellcor™ sensor from the sensor port.
Table 6-4.
Possible Data Interface Settings
Option
Possible Setting
Factory Default
Remote settings
Wireless LAN: ASCII, SPDOut
Disconnected
LAN: ASCII, SPDOut
Disconnected
Nurse call priority, RS-232
No, normally -
No
Serial connection
ASCII: 9600 or 19200 baud
ASCII, 9600 baud
Clinical: 19200 baud
SPDout: 19200 or 115200 baud
Philips: 19200 baud
OFF
WLAN settings
Wi-Fi or Network Connection
None
LAN settings
Table 6-5.
Possible Service Functions
Option
Possible Setting
Factory Default
Service Options
Touch screen calibration
Yes, No
No
Log Access
View error, event, or software log
Requires USB flash drive
Export error, event, or software log
Clear logs
Requires confirmation of desire to clear logs; not reversible
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......