Performance Tests
Service Manual
10-29
SpO
2
Test
To perform the SpO
2
Test
1.
Ensure the monitoring system is connected to an AC power source.
2.
Connect a DOC-10 interface cable to the sensor port.
3.
Turn on the monitoring system by pressing the POWER ON key.
4.
Verify the word PLETH appears at the top left corner of the monitoring screen.
5.
Connect the SRC-MAX tester to the interface cable.
6.
Verify the listed system behaviors.
Figure 10-8.
SpO
2
Test: SpO
2
75, BPM 60
a.
Active visual and audio alarms.
b.
Saturation indication between 73 and 77 inclusive, creating an SpO
2
LOW
alarm condition.
c.
BPM indication between 57 and 63 inclusive.
d.
Pulse waveform of approximately 1/2-inch peak to peak (P-T-P) amplitude.
Actual amplitude may vary but will be a reference for low pulse amplitude/low
light patients.
7.
Press %SpO
2
.
8.
Verify the 90% LED lights. The monitoring system registers SpO
2
increases and
stabilizes to a value between 88 to 92 inclusive.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......