Chassis Disassembly and Reassembly
Service Manual
12-19
9.
Ensure both halves remain on a static-free work surface.
10.
Review the following table for identifying Main PCB connectors. The table identi-
fies cables and connectors beginning on the top, right hand corner of the
Main PCB and proceeding clockwise. Some cables are barely visible or require
special routing.
11.
Disconnect the spade connector (GND) from the J21 mating blade on the Main
PCB.
12.
Disconnect the five-pin battery connector from the J23 receptacle on the Main PCB.
13.
Disconnect the four-pin power connector from the J22 receptacle on the Main PCB.
14.
Disconnect the three-pin fan connector from the J2 receptacle on the Main PCB.
Table 12-3.
Main PCB Connections
Receptacle
Cable Description
Cable Color
J5
Two-pin speaker cable
Red and black
J7
Five-connector flex cable
Translucent aqua
J15
Main ribbon cable
Gold
J2
Three-pin fan cable
Red, black, and yellow
J22
Four-pin power cable
Red and black
J23
Five-pin battery cable
Black
J21
Spade
Green GND
J10
Six-connector flex cable
Gold
J6
Four-connector flex cable
Gold
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......