Setting Institutional Defaults
Service Manual
10-7
To import master institutional defaults
Note:
Settings may only be imported from exports created with the same software version.
Do not attempt to import settings from a monitoring system with a different software
version.
Note:
Network settings are not affected by the import.
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select IMPORT AND EXPORT SETTINGS.
4.
Select IMPORT SETTINGS.
5.
Insert a USB flash drive containing the master institutional default settings.
6.
Press NEXT.
7.
Press NEXT after ensuring the
ApplicationSettingsFile.xml
option is selected.
8.
Remove the USB flash drive at the prompt.
9.
Press FINISH.
10.
Press EXIT MENU.
11.
Press EXIT AND RESTART.
12.
Press YES.
To restore factory defaults
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select RESTORE FACTORY DEFAULTS.
4.
Press YES.
5.
Press FINISH.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......